DIAGNOSTICS
10DEC05
22005 - MSC BOARD A1: Error during download of MSC Software
DG4825-1
Page
Description: The boot of the MOTION SYSTEM CONTROL received a signal that is
105 of 180
Cause:
Action:
22006 - MSC BOARD A1: Error during download of MSC Software
Description: The receive buffer of the boot program received excessive data.
Cause:
Action:
22007 - MSC BOARD A1: Error during download of MSC Software
Cause:
Action:
22008 - MSC BOARD A1: Error during download of MSC Software
Description: The boot program for the MOTION SYSTEM CONTROL BOARD A1
not correct during the downloading of the software to the MOTION
SYSTEM CONTROL BOARD A1.
The software did not download.
1.
Download the software again.
2.
Do the checkout for the
3.
If this error continues, install a new MOTION SYSTEM CONTROL BOARD A1.
The software did not download.
1.
Download the software again.
2.
Do the checkout for the
3.
If this error continues, install a new MOTION SYSTEM CONTROL BOARD A1.
The software did not download.
1.
Download the software again.
2.
Do the checkout for the
started to send a message on an RS-232 channel that is not correct.
MOTION SYSTEM CONTROL BOARD
MOTION SYSTEM CONTROL BOARD
MOTION SYSTEM CONTROL BOARD
Error Codes
A1.
A1.
A1.