Chapter 10 Pass/Fail Test

Cursor Region Mode

In the cursor region mode, you can set the test mask range by adjusting the
positions of the cursors (cursor A and cursor B).
Press Utility  Pass/Fail  Range to open the mask range menu. In this menu,
1. Specify the test mask range to the cursor region
2. Adjust the cursor positions (used to determine the test mask range)
3. Adjust the horizontal tolerance range
MSO4000/DS4000 User's Guide
800.544.2843
Press Range to switch to "Cursor".
At this point, two gray cursor lines (used to indicate the current test mask range)
are displayed on the screen. Cursor A is on the left and cursor B is on the right.
Press CursorA and turn the multi-function knob
horizontal position of cursor A (namely adjust the left margin of the test
mask range).
Press CursorB and turn the multi-function knob
horizontal position of cursor B (namely adjust the right margin of the test
mask range).
Press CursorAB and turn the multi-function knob
horizontal positions of cursor A and cursor B at the same time (namely
adjust the left margin and right margin of the test mask range). At this point,
the horizontal distance between cursor A and cursor B remains unchanged.
Note:
The horizontal distance between cursor A and cursor B should be at least 1
grid.
When CursorA, CursorB or CursorAB is currently selected, pressing the
multi-function knob
Press X Mask and turn the multi-function knob
tolerance range of the test mask. The adjustable range is from 0.02 div to 4.00
div. During the adjustment, two white curves would be displayed in the interface
to show the outline of the current mask.