Chapter 3 Horizontal System Setting
Application example: measure the phase deviation of the input
signals of two channels.
Method 1: Use Lissajous method
1.
2.
3.
4.
Note: The maximum sample rate of the oscilloscope in the X-Y mode is 2.0 GSa/s.
Generally, longer sample waveform ensures better figure; but, as being limited by
the memory depth, longer waveform will require a reduction in the sample rate
(please refer to "To Set the Memory Depth"). Therefore, in this measurement
process, reducing the sample rate appropriately can improve the displayed Lissajous
figure.
MSO4000/DS4000 User's Guide
800.544.2843
Connect a sine signal to CH1 and then connect a sine signal with the same
frequency and amplitude but a 90° phase deviation to CH2.
Press AUTO and turn on the X-Y time base mode.
If the XY1 area display is turned on, please move to the next step. If the XY1
area display is turned off, press XY1 to switch to "On".
Use VERTICAL
POSITION of CH1 and CH2 to display the signals at the
center of the screen and use VERTICAL
the signals easy to observe (if necessary, please refer to "To Adjust the
Vertical Scale" to switch the vertical scale adjustment mode). At this point, the
circle as shown in the figure below should be displayed.
According to the phase deviation measurement principle diagram and the
measurement result in the figure above, it can be calculated that
Therefore, the phase deviation angle
www.calcert.com
15
10
20
5
25
0
30
SCALE of CH1 and CH2 to make
θ
.
=
±
arcsin
1
=
±
90
°
RIGOL
A
C
.
=
1
B
D
3-7