Rigol MSO4000 Series User Manual page 192

Digital oscilloscope
Hide thumbs Also See for MSO4000 Series:
Table of Contents

Advertisement

RIGOL
Delay and Phase Parameters
Source A
Source B
1. Dly A →B : the time difference (delay) between the threshold middle values
of the rising edges of source A and source B. Negative delay indicates that the
rising edge of source A occurred after the rising edge of source B.
2. Dly A →B : the time difference (delay) between the threshold middle values
of the falling edges of source A and source B. Negative delay indicates that the
falling edge of source A occurred after the falling edge of source B.
3. Dly A →B : the time difference (delay) between the threshold middle value
of the rising edge of source A and the threshold middle value of the falling edge
of source B. Negative delay indicates that the rising edge of source A occurred
after the falling edge of source B.
4. Dly A →B : the time difference (delay) between the threshold middle values
of the falling edge of source A and the threshold middle value of the rising edge
of source B. Negative delay indicates that the falling edge of source A occurred
after the rising edge of source B.
5. Phs A →B : the phase difference (expressed in degree) between the
threshold middle values of the rising edges of source A and source B. The
operation formula is:
Wherein,
A →B ;
6. Phs A →B : the phase difference (expressed in degree) between the
threshold middle values of the falling edges of source A and source B. The
6-26
800.544.2843
Period
Delay
PhaseA
B
R
denotes phase A →B ;
PhaseA
B
R
R
denotes the period of source A.
Period
sourceA
www.calcert.com
15
10
20
5
25
0
30
Chapter 6 Operations and Measurements
DelayA
B
R
R
=
×
360
R
Period
sourceA
DelayA
MSO4000/DS4000 User's Guide
°
denotes delay
B
R
R
sales@calcert.co