Rigol MSO4000 Series User Manual page 220

Digital oscilloscope
Hide thumbs Also See for MSO4000 Series:
Table of Contents

Advertisement

RIGOL
respectively.
XY2: the X cursors and Y cursors are used to measure the amplitude values
of the specified points (cursor crossing points) of CH3 and CH4 waveforms
respectively.
3. Ajust the X/Y cursor positions
1) Ajust the X cursor positions
2) Ajust the Y cursor positions
Note: When any of CursorA, CursorB and CursorAB is currently selected,
you can press the multi-function knob
the three items.
6-54
800.544.2843
Press Type to select "X".
Press CursorA and turn the multi-function knob
horizontal position of cursor A (X cursor). The adjustable range is
limited within the screen.
Press CursorB and turn the multi-function knob
horizontal position of cursor B (X cursor). The adjustable range is
limited within the screen.
Press CursorAB and turn the multi-function knob
horizontal positions of cursors A and B (X cursors) at the same time.
The adjustable range is limited within the screen. The horizontal
distance between cursors A and B (X cursors) keeps unchanged.
Press Type to select "Y".
Press CursorA and turn the multi-function knob
vertical position of cursor A (Y cursor). The adjustable range is limited
within the screen.
Press CursorB and turn the multi-function knob
vertical position of cursor B (Y cursor). The adjustable range is limited
within the screen.
Press CursorAB and turn the multi-function knob
vertical positions of cursors A and B (Y cursors) at the same time. The
adjustable range is limited within the screen. The vertical distance
between cursors A and B (Y cursors) keeps unchanged.
www.calcert.com
15
10
20
5
25
0
30
Chapter 6 Operations and Measurements
to adjust the
to adjust the
to adjust the
to adjust the
continuously to quickly switch among
MSO4000/DS4000 User's Guide
to adjust the