Chapter 14 System Function Setting

Aux Output

Users can set the oscilloscope to output the specified type of signal from the [Trig
Out/Calibration] connector at the rear panel.
Press Utility  AuxOutput and turn the multi-function knob
desired signal type. You can also press AuxOutput continuously to switch the
current signal type.
Note: You can also modify the signal type of AuxOutput in the pass/fail test
function (refer to "To Set the Output Form of Test Failure").
MSO4000/DS4000 User's Guide
800.544.2843
TrigOut
The oscilloscope outputs a signal that can reflect the current capture rate of the
oscilloscope from the [Trig Out/Calibration] connector at the rear panel at
each trigger. Connect the signal to a waveform display device and measure the
frequency of the signal; the measurement result is the same as the current
capture rate.
Fast
The oscilloscope outputs a fast edge signal with 500 ps rise time from the [Trig
Out/Calibration] connector at the rear panel. The signal can be used in the
self-calibration of the oscilloscope (refer to "Self-calibration").
GND
The oscilloscope outputs a ground level from the [Trig Out/Calibration]
connector at the rear panel.
PassFail
The oscilloscope will output a pulse from the [Trig Out/Calibration]
connector at the rear panel when failed waveforms are detected. This signal can
be connected to other control systems to conveniently view the test results.