Chapter 6 Operations and Measurements
4. Specify the display area
5. Specify the horizontal (X axis)/vertical (Y axis) unit
MSO4000/DS4000 User's Guide
800.544.2843
You can select measurement items of the X cursors and Y curosrs by
setting the "horizontal measurement unit" (Time Unit) and
The measurement results include A->X, B->X, A->Y, B->Y, △X, 1/△X
"vertical measurement unit" (Vertical Unit) respectively.
(only displayed when Time Unit is set to "s" or "Hz") and △Y.
When the delayed sweep is turned off (pressing HORIZONTAL
can turn the delayed sweep on or off), Screen is set to "Main"
automatically and cannot be modified.
When the delayed sweep is turned on (pressing HORIZONTAL
can turn the delayed sweep on or off), the screen is divided into the Main
and Zoom areas.
Press Screen to select "Main" or "Zoom".
When "Main" is selected, the cursors are displayed in the Main area and
are used to measure the parameters in the Main area; the
measurement results are displayed in the Zoom area.
When "Zoom" is selected, the cursors are displayed in the Zoom area
and are used to measure the parameters in the Zoom zrea; the
measurement results are displayed in the Main area.
Note: When the measurement source is set to MATH, Screen is set to
"Zoom" automatically and cannot be modified.
1)
When the cursor display type is "X" or "X-Y", you can select the horizontal
measurement unit.
Press Time Unit and turn the multi-function knob
horizontal measurement unit. You can also press Time Unit continuously
to switch the current horizontal measurement unit. The units available
include "s", "Hz", "°" or "%".
reference). The measurement results include A->X, B->X, △X and
s: measure the time values at the X cursors (take the trigger position as
1/△X. Wherein, A->X, B->X and △X denote time and 1/△X denotes
frequency.
Hz: measure the frequency values at the X cursors (take the trigger
www.calcert.com