Rigol MSO4000 Series User Manual page 235

Digital oscilloscope
Hide thumbs Also See for MSO4000 Series:
Table of Contents

Advertisement

Chapter 8 Protocol Decoding
3. Set the data line
4. Adjust the analog channel threshold level
Channel Signal Amplitude
MSO4000/DS4000 User's Guide
800.544.2843
2) Specify the edge type of sampling
Press Slope and turn the multi-function knob
sample the channel data on the rising edge (
rising&falling edges (
continuously to switch the current edge type.
1) Set the bus bits
Press Bus Bits and turn the multi-function knob
of the parallel bus (namely the number of bits of each frame of data). The
range is from 1 to 20 and the default is 8 (Bit0, Bit1, Bit2...).
2) Specify data channel for each bit
First, press CurrentBit and turn the multi-function knob
bit that needs to specify a channel. The range available is from 0 to (the
current bus bits – 1). By default, bit 0 is selected. For example, when the
bus bits is 20, the range available is from 0 to 19.
Then, press Channel, turn the multi-function knob
desired channel and press the knob. You can also press Channel
continuously to switch the channel currently selected. The channels
available include D0 to D15 and CH1 to CH4.
When the clock channel or data channel selected include analog channels (CH1
to CH4), you can set the threshold level of the specified analog channels
according to your need. When the channel signal amplitude is greater than the
preset threshold, it is judged as logic "1"; otherwise logic "0".
Channel Logic Level
www.calcert.com
15
10
20
5
25
0
30
) of the clock signal. You can also press Slope
1
0
0
to set the oscilloscope to
), falling edge (
to set the data width
to select the
to select the
Threshold Level
sales@calcert.co
RIGOL
) or
8-3

Advertisement