Measurements
selected measurement item that you've added; click or tap Remove All to delete
all the displayed measurement items.
In the "Result" sidebar, select a measurement item and drag it to the right to
delete it quickly
10.3

Cursor Measurements

Cursor measurement can measure the X axis values (e.g. Time) and Y axis values (e.g.
Voltage) of the selected waveform. Before making cursor measurements, connect the
signal to the oscilloscope to acquire stable display. The cursor measurement function
provides the following two cursors.
X Cursor
X cursor is a vertical solid/dotted line that is used to make horizontal
adjustments. It can be used to measure time (s) and frequency (Hz).
-
-
Y Cursor
Y cursor is a horizontal solid/dotted line that is used to make vertical
adjustments. It can be used to measure amplitude (the unit is the same as that of
the source channel amplitude).
DHO800 User Guide
140

Figure 10.8 Cursors

Cursor A is a vertical solid line (
and Cursor B is a vertical dotted line (
screen).
In the XY cursor mode, X cursor is used to measure the waveform
amplitude of Source X.
is displayed at the bottom of the screen),
is displayed at the bottom of the