Chapter 5 Trigger
4. Specify the trigger mode
5. Set the trigger holdoff and noise rejection
MSO4000/DS4000 User's Guide
800.544.2843
adjust the pulse width within a relatively larger range, this function can help
you quickly finish the setting.
When the pulse width condition is set to
press Upper Limit and Lower Limit respectively and use the
multi-function knob
lower limit of pulse width. The range of the upper limit is from 12 ns to 4 s.
The range of the lower limit is from 4 ns to 3.99 s.
Note:
The lower limit of the pulse width must be lower than the upper limit.
You can use the outer knob of the navigation knob to quickly adjust the
upper limit and lower limit of pulse width within a relatively larger
range.
Press Sweep to select "Auto", "Normal" or "Single" (please refer to "Trigger
Mode"). The corresponding status light of the trigger mode currently selected in
the trigger control area (TRIGGER) at the front panel turns on.
Press Setting to enter the trigger setting submenu. In this submenu:
Press Holdoff to adjust the holdoff time (please refer to "Trigger
1)
Holdoff"). The range is from 100 ns to 10 s.
Press Noise Reject to turn the noise rejection on or off (please refer to
2)
"Noise Rejection").
Note: When the trigger source is D0 to D15, the noise rejection is invalid.