To Set The Delay Calibration Of The Digital Channels - Rigol MSO4000 Series User Manual

Digital oscilloscope
Hide thumbs Also See for MSO4000 Series:
Table of Contents

Advertisement

Chapter 7 Logic Analyzer

To Set the Delay Calibration of the Digital Channels

When using an oscilloscope for actual measurement, the transmission delay of the
probe cable may bring greater error (zero offset). Zero offset is defined as the offset
of the crossing point of the waveform and threshold level line relative to the trigger
position (refer to "To Set the Delay Calibration"). MSO4000/DS4000 allows users
to set a delay time to calibrate the zero offset of the corresponding channel.
Press LA  Delay-Cal and turn the multi-function knob
calibration time at the specified step and the range available is from -100 ns to 100
ns. Pressing the multi-function knob
Note: The delay calibration time setting is related to the instrument model and the
current horizontal time base. The larger the horizontal time base is, the larger the
setting step will be. Take MSO4054 as an example; when the horizontal time base is
50 ns, the step is 1 ns; when the horizontal time base is 2 μs, the step is 40 ns; when
the horizontal time base is greaten than or equal to 10 μs, the delay calibration time
cannot be adjusted.
MSO4000/DS4000 User's Guide
800.544.2843
can quickly reset the delay time to 0.00 s.
www.calcert.com
15
10
20
5
25
0
30
RIGOL
to adjust the delay
7-11
sales@calcert.co