Chapter 10 Pass/Fail Test

To Turn the Test Result Display On/Off

Press Utility  Pass/Fail  Enable and select "ON" to enable the pass/fail test
function; then, press Message to turn on or off the display of the test result statistic
information.
The test result statistic information includes the number of failed frames (Fail), the
number of passed frames (Pass) and the total number of frames (Total). The
information (as shown in the figure below) is displayed at the upper-right side of the
screen.

To Set the Output Form of Test Failure

Press Utility  Pass/Fail to open the pass/fail test menu. In this menu, you can
set the operation that the oscilloscope will execute when failed waveform is detected
according to your need.
1. Turn the stop on fail on or off
MSO4000/DS4000 User's Guide
800.544.2843
Press StopOnFail to select "ON" or "OFF".
ON: when failed waveform is detected, the oscilloscope will stop the test
and enter the "STOP" state.
At this point, the test result statistic information of the last test remains the
same on the screen if Message is set to "ON"; a pulse will be output from
the [Trig Out/Calibration] connector at the rear panel if AuxOutput is
set to "ON".
OFF: the oscilloscope will continue the test even though failed waveforms
are detected.
At this point, the test result statistic information on the screen will update
continuously if Message is set to "ON"; a pulse will be output from the