Screen Region Mode

In the screen region mode, the test mask range is the whole careen region.
Press Utility  Pass/Fail  Range to open the mask range menu. In this menu,
1. Specify the test mask range to the screen region
Press Range to switch to "Screen".
2. Adjust the horizontal tolerance range
Press X Mask and turn the multi-function knob
tolerance range of the test mask. The adjustable range is from 0.02 div to 4.00
div. During the adjustment, two white curves would be displayed in the interface
to show the outline of the current mask.
3. Adjust the vertical tolerance range
Press Y Mask and turn the multi-function knob
tolerance range of the test mask. The adjustable range is from 0.03 div to 4.00
div. During the adjustment, two white curves would be displayed in the interface
to show the outline of the current mask.
4. Create the test mask
Press Create to immediately creat the test mask according to the current
settings (the test mask range, the horizontal tolerance range and the vertical
tolerance range) and display the current test mask area (the area not covered by
blue within the screen) on the screen, as shown in the figure below.
10-4
800.544.2843
Figure 10-1 Test Mask (Screen Region)
www.calcert.com
15
10
20
5
25
0
30
Chapter 10 Pass/Fail Test
to adjust the horizontal
to adjust the vertical
MSO4000/DS4000 User's Guide