To Turn the Measurement History On/Off

This oscilloscope allows users to view the measurement history (only the
measurement items of which the measurement results are currently displayed) of
one-key measurement.
Press Measure  History to open the measurement history menu. In this menu,
press History to turn on or off the measurement history.
Note: When no measured item is currently displayed, the measurement history
cannot be turned on.
Press Display to select "Graph" or "Table".
Graph: display the measurement results from multiple measurements of at most
5 measurement items that are turned on last in graph mode. The measurement
points are connected using linear interpolation.
Table: display the last 10 measurement results of at most 5 measurement items
that are turned on last in table form.