RIGOL
information. The line currently selected is highlighted with brown shading
and the decoded data of this line is dispalyed at the lower-left side of the
event table.
Note:
2)
If a USB storage device is currently connected to the instrument, you can
press Export to export the decoded data list to the external USB storage
device in CSV format.
8. Address information during decoding
In I2C bus, each frame of data starts with the address information (include the
write address and read address). This oscilloscope uses blue patch to represent
the address ID. In the address ID, "Read" denotes the read address (for
example,
Note: You can also set the address information to include or not include the
"R/W" bit (refer to "Include R/W").
8-18
800.544.2843
The display format of the decoded data in the event table is determined
by the bus display format currently selected (refer to "Format").
When adjusting the horizontal time base, the waveform displayed on
the screen will change and the total number of lines of decoding
information in the event table will change.
The decoded data information displayed in the bus is related to the
horizontal time base. You can view the detailed information by reducing
the horizontal time base.
) and "Write" denotes the write address (for example,
).