6.4 System operating temperature configuration

The system operating temperature configuration can be adjusted to suit process
demands. Contact your supplier for instructions on how to access the configuration
menu (through the PDT) and for process specific recommendations.
Use the following procedure:
1.
Access the Config menu on the PDT (requires an access code, contact supplier).
2.
Scroll to Temp Set-points - press Enter.
3.
Scroll to Set DP Temp - press Enter.
4.
Scroll to desired temperature - Press Enter.
Note:
The Dry Pump set-point is the desired surface temperature of the internal low
vacuum stage.
Refer to
settings and ranges.
Table 32 Dry pump and booster temperature settings
Temperature Setting
Dry pump
(Standard variants)
iXM600, iXM1200
Dry pump
(Standard variants)
iXM200, iXM3000
Dry pump
(T and TX variants)
Booster
(Only TX variants)
M56635880_D
M56635880_D - Operation
Table: Dry pump and booster temperature settings
Range in which temperature can be
set(°C)
60 - 110
60 - 110
60 - 130
-
Page 65
for default temperature
Default value (°C)
80
110
130
85
10/2020 - ©Edwards Limited