Connect The Accessories - Edwards iXM Series Instruction Manual

Dry pump system
Table of Contents

Advertisement

1.
Use a 3/8 inch BSP male pipe fittings to attach the female quick-release connector
to the cooling water supply hose.
2.
Use a 3/8 inch BSP female pipe fitting to attach the male quick-release connector
to the cooling water return hose.
3.
Remove the dust-caps from the cooling water inlet and outlet.
4.
Connect the water return hose to the cooling water outlet
connectors on the rear of the
supply line close to the pump and connect the water supply hose to the cooling
water inlet
5.
Turn on the cooling water supply.
6.
Switch on the electrical power to the system. Wait for approximately 2 minutes.
Use the Pump Display Terminal accessory (PDT) refer to
(PDT)
Test Mode option. Select Water On and press Enter. This command will open all the
main pump water valves for approximately 30 seconds.
7.
While the main pump water valves are open, adjust the water flow rate so that it
meets the requirements given in
required to adjust the water flow rate, access the Test Mode option again and
select Disabled. Press Enter. Access the Test Mode option one more time and
select Water On same as before. Repeat this test until the water flow rate has been
set to the required value.
8.
Once the cooling water flow rate has been set, go to the Setup menu on the PDT.
Select the Test Mode option. Select the option Disabled and press Enter. The valves
should now close and the flow rate displayed by the water flow meter will
decrease.
9.
Switch off the electrical power to the system.
10.
The water flow meter can now be removed.
11.
Inspect the water hoses, pipelines and connections and check that there are no
leaks.
12.
Turn off the water supply

4.6.10 Connect the accessories

WARNING: INSTALLATION SAFETY
Risk of injury. When you install accessories in the dry pumping system enclosure,
make sure that the pump is switched off and Lockout and Tagout the electrical supply
before removing the enclosure panels.
WARNING: HOT SURFACES
Risk of burn injury. The surfaces of the dry pump, booster and spools are very hot
when the system is running. Let these surfaces cool to safe temperatures before you
install accessories in the system enclosure.
M56635880_D
M56635880_D - Installation
(Figure: Controls/connectors on the rear of the
on page 93 and press the Setup button. Under the Setup menu select the
Page 47
system). Install a water flow meter in the water
Table: Water cooling system
(Figure: Controls/
system).
Pump Display Terminal
data. If more time is
10/2020 - ©Edwards Limited

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents