Edwards iXM Series Instruction Manual page 77

Dry pump system
Table of Contents

Advertisement

Fault
Warning 152.01 - Valve Not Shut
Cause
The foreline gate valve has failed to close.
Remedy
Examine the wiring and air supply to the Gate valve. This warning will only clear
when the valve has successfully been opened and closed.
Fault
Warning 153.01 - Valve Not Open
Cause
The foreline gate valve has failed to open.
Remedy
Examine the wiring and air supply to the Gate valve. This warning will only clear
when the valve has successfully been closed and opened.
Fault
Warning 161.01 - GV Missing
Cause
The foreline gate valve is disconnected or missing (open circuit)
1. Make sure that the sensor is installed and is correctly connected.
Remedy
2. If this warning continues, contact us.
Fault
Warning 161.01 - GV Fault
Cause
The foreline gate valve is faulty (short circuit)
1. Make sure that the sensor is installed and is correctly connected.
Remedy
2. If this warning continues, contact supplier.
Fault
Warning 176.01 - MB zzzzzzzzzzzzz
Cause
Booster Inverter gives warning code xxxx yyyy / Diagnostic registers aaaa bbbb.
Remedy
Refer to
Fault
Warning 176.13 - No MB Inv Comms
Cause
The pump controller cannot communicate with the booster inverter.
Remedy
Examine the wiring between the pump controller and the booster inverter.
Fault
Warning 186.01 - DP zzzzzzzzzzzzz
Cause
DP Inverter gives warning code xxxx yyyy/Diagnostic registers aaaa bbbb.
Remedy
Refer to
Fault
Warning 186.13 - No DP Inv Comms
Cause
The pump controller cannot communicate with the dry pump inverter.
Remedy
Examine the wiring between the pump controller and the dry pump inverter.
M56635880_D
M56635880_D - Fault finding
Inverter warnings and alarms
Inverter warnings and alarms
Page 74
on page 83 for more information
on page 83 for more information
10/2020 - ©Edwards Limited