Edwards iXM Series Instruction Manual page 23

Dry pump system
Table of Contents

Advertisement

Description
▪ for 380 - 460 V a.c. systems
* Typical earth leakage values were measured at steady-state conditions.
Note that higher leakage currents can occur:
i) in transient conditions such as power on or pump acceleration
ii) with abnormal supply configurations such as a missing or earthed phase or
unbalanced supply voltages.
Contact us for more information about configuration requirements for earth leakage
reduction.
Table 9 Electrical connections
Description
Mains connection
Refer to
Installation
on page 25
for wiring diagram
PDT interface (front)
System interface (rear)
Ethernet interface
USB flash memory stick/drive
EMS interface
External controlled stop button
Pin 1 supply, Pin 2 - return
Internal controlled stop button§
M56635880_D
M56635880_D - Technical data
Units
mA
Mating connector description / external
supply rating
Refer to
Table: Pump electrical data
mains connector installed to each variant.
Insert type is one of:
Harting Han 40A, axial module, part num-
ber: 09 14 002 2701, 2.5 – 8 mm2 fine stran-
ded wire, Hood part number: 09 30 006
0442 or
Insert type is : Weidmüller 1789980000, 2.5
– 8 mm2 fine stranded wire, Hood part
number : 1652680000 (VDE 0295 class 5, re-
fer to
Table: Wire assembly according to
VDE
0295), 8.9 mm maximum insulation di-
ameter or
Han K 4/4-F finger safe 09 38 008 2703, 6-16
2
mm
fine stranded wire (VDE 0295 class 5,
refer to
Table: Wire assembly according to
VDE
0295), 8.9 mm maximum insulation di-
ameter
XLR type 5-way plug
XLR type 5-way plug
Standard RJ45 type or Neutrik® EtherCon®
RJ45
USB Type A flash stick only (no cable con-
nection allowed)
XLR type 6-way plug
Page 20
Rating
< 10
Internal supply rating
for the
(IEEE802.3i 10 Base T
24 V d.c. 100 mA
10/2020 - ©Edwards Limited
24 V d.c. 0.2 A
*
24 V d.c. 0.75 A
Ethernet)
5 V d.c. 100 mA

Hide quick links: