Heater Data - Edwards iXM Series Instruction Manual

Dry pump system
Table of Contents

Advertisement

Table 11 Water cooling system data
Description
Maximum supply pressure
Necessary flow rate
Maximum allowable system
differential pressure
Minimum necessary pres-
sure differential across sup-
ply and return
Supply temperature range
Water type
Maximum particle size
Acidity
Hardness (< 100 mg of Ca-
CO
per litre)
3
Resistivity
Materials in contact with
cooling water
Water inlet connection
Water outlet connection
* For iXM1200 TX variants, if continuous operation is required with 60 slm (Air) or more
at the pumped gas inlet, 5 l/min is recommended. Please contact us for more
information.

3.8 Heater data

Table 12 TX variant heater data
Description
Exhaust heater control temperature
Exhaust heater power consumption
iXM1200TX
Typical booster heater control temperature
Typical booster heater power consumption
iXM1200TX
M56635880_D
M56635880_D - Technical data
Units
barg
psig
litres/min
bar
bar
°C
2
mm
pH
ppm of CaCO
3
ohm-cm
Stainless steel
and fluoroelas-
Page 22
1200
3000
600/600T
7
4
0.75
0.75
5 to 30
Treated or non-corrosive industrial
0.03
6.5 to 8.0
< 100
1 k £ r £ 1000 k
Stainless steel, Nitrile, PTFE and fluoroe-
tomer
3/8 inch BSP male quick connector
3/8 inch BSP female quick connector
Units
°C
W
°C
kW
1200T
1200TX
6.9
100
*
4 (5
)
5.5
1.0
lastomer
Rating
160
110
85
1.2
10/2020 - ©Edwards Limited
200
200T
3
0.75

Hide quick links:

Advertisement

Table of Contents