1. Examine the equipment recently connected to the pump. Make sure that is
Remedy
installed and correctly connected.
2. If the warning does not disappear power cycle the pump.
3. If the warning re-appears after power cycling, contact us.
Fault
Event 258.63 - USB V oload
Cause
Faulty equipment has just been plugged in
1. Examine the equipment recently connected to the pump. Make sure that is
Remedy
installed and correctly connected.
2. If the warning does not disappear power cycle the pump.
3. If the warning re-appears after power cycling, contact us.

8.5 Inverter warnings and alarms

The inverters used to drive the dry pump and booster can also give warnings and alarms
if problems are identified. Inverter warning and alarm codes are displayed by the PDT if
installed.
Table 38 Inverter warning and alarm status registers
Inverter warning
Text displayed on
and alarm status
PDT as alert and fault
registers
history
0123456789012
0800 0000
NO_INV_PARAMS
0000 0010
POWER_DOWN___
0000 0008
POWER_DOWN___
0040 0000
EMS_ACTIVATED
0080 0000
MISSING_PHASE
0000 0200
MTR_TEMP_HIGH
0000 0020
MTR_TEMP_HIGH
0000 0400
INV_TEMP_HIGH
0000 0040
INV_TEMP_HIGH
0000 0800
INV_TEMP_HIGH
0008 0000
INV_TEMP_HIGH
0000 0100
MISSING_PHASE
8000 0000
LOW_SPEED____
M56635880_D
M56635880_D - Fault finding
Description
No inverter parameters set loa-
ded.
Supply voltage collapsed (power
off)
Supply voltage collapsed (power
off)
EMS activated
Missing Phase in mains supply
High Motor Temperature
High Motor Temperature
High Inverter Sink Temperature
High Inverter Sink Temperature
High Inverter Processor Temper-
ature
High Inverter Temperature Trip
Missing Phase in mains supply
Acceleration timeout (Pump
Seized/overload)
Page 83
Recommended action
No inverter parameters
set loaded.
Check Supply Voltage
Check Supply Voltage
EMS activated
Missing Phase in mains
supply
Check cooling
Check cooling
Check cooling
Check cooling
Check cooling
Check cooling
Missing Phase in mains
supply
Acceleration timeout
(Pump Seized/overload)