Figure 10 Connect the pump inlet (iXM600 and iXM1200)

1. Remove screws
1. Remove screws
Table 17 Half claw clamps
Flange diameter
ISO63
ISO100
ISO200

4.6.3 Connect the pump exhaust

Note:
The system is supplied with a trapped O
expected exhaust temperatures. Make sure that the seals are always replaced with the
same type.
Use the trapped O-ring seal and clamp supplied to connect the exhaust outlet on the
exhaust pipe, to the exhaust extraction system. Refer to
the rear of the system

4.6.4 Connect to the factory extraction system (optional)

An extraction port is given to allow secondary exhaust ventilation. When connected, the
airflow operates continuously and will extract substances of concern that can be
released during a failure.
Refer to
Accessories
Refer to
M56635880_D
M56635880_D - Installation
Half claw clamp
part number
C10007093
C10007093
C10011093
Tracer gas analysis
on page 23 for the necessary extraction rates. Refer to
on page 93 for the necessary exhaust extraction cover kit.
Figure: Front view of the system
2. Remove cover
2. Remove cover
Quantity required
‑ ring or a metal seal based on the pump type and
for location of the extraction port.
Page 34
Tightening torque
(Nm)
4
8
12
Figure: Controls/connectors on
10/2020 - ©Edwards Limited
5
5
5