Inverter warning
Text displayed on
and alarm status
PDT as alert and fault
registers
history
4000 0000
LOW_SPEED____
0004 0000
CURRENT_HIGH_
0000 0001
MTR_TEMP_LOW_
0000 0002
INV_TEMP_LOW_
0010 0000
INV_TEMP_LOW_
0002 0000
INV_VOLTS_HI_
0020 0000
INV_OUT_SHORT
0001 0000
SPEED_HIGH___
0000 4000
MTR_OVERLOAD_
A:1000 W:0000
INV_COMMSLOSS
D1:8000 D2:0000
1000 0000
INV_TEST_FLT
0000 8000
INV_TEST_WRN
2000 0000
INV_TEST_FLT
A:0200 W:0000
INV_TEST_FLT
0100 0000
INV_PWM_TRIP_
A0000 W0000
INV_SW_UPGRDE
D1:0000 D2:0000
Ss1:0000 Ss2:0020
Anything else
INV_OTHER_FLT
M56635880_D
M56635880_D - Fault finding
Description
Overload timeout (Pump Seized/
overload)
Over Current Trip (Possible Met-
al-metal contact)
Low Motor Temperature
Low Inverter Sink Temperature
Low Inverter Temperature trip
Check supply voltage
Check for windmilling
Inverter Output Short Circuit -
hard to fix in the field - possibly
requires inverter change
Internal Inverter Control fault
Motor overload
(Pump Seized/overload)
Loss of communications be-
tween drives in dual inverter
Inverter self-test fault
Inverter self-test warning
Inverter self-test fault
Inverter self-test fault
All other sources (and causes) of
PWM Trip inserted before PWM
trip in the decode sequence
Software upgrade has just occur-
red - not displayed on PDT as
alert - only appears in fault log
Inverter unidentified fault
Page 84
Recommended action
Overload timeout
(Pump Seized/overload)
Over Current Trip (Pos-
sible Metal-metal con-
tact)
Low Motor Tempera-
ture
Low Inverter Sink Tem-
perature
Low Inverter Tempera-
ture trip
Check supply voltage
Check for windmilling
Inverter Output Short
Circuit
Internal Inverter Con-
trol fault
Motor overload
(Pump Seized/overload)
Loss of communications
between drives in dual
inverter
Inverter self-test fault -
swap inverter
Inverter self-test warn-
ing (no need to swap in-
verter)
Inverter self-test fault -
swap inverter
Inverter self-test fault -
swap inverter
Inverter trip occurred.
Report to us.
Software upgrade has
just occurred - not dis-
played on PDT as alert -
only appears in fault log
Report to us.