Edwards iXM Series Instruction Manual page 41

Dry pump system
Table of Contents

Advertisement

WARNING: HAZARDOUS VOLTAGES
Risk of electric shock. All connections to the interface control must be double
insulated or have equivalent protection. Do not connect voltages greater than
30 V a.c. or 60 V d.c. to the control/interface connections.
WARNING: HAZARDOUS VOLTAGES
Risk of electric shock. The power wiring to the system must be properly protected
and the routing must be away from possible hazards.
WARNING: HOT SURFACE
Risk of burn injury. The exhaust of the dry pumping system will heat up when the
pump is in operation.
CAUTION: ELECTRICAL SAFETY
Risk of damage to equipment. Do not connect voltages greater than specified in this
manual to the control/interface connections as damage can be caused to the
interface control.
WARNING: HAZARDOUS VOLTAGES
Risk of electric shock. The secondary protective earth (ground) is required in case of
failure of the primary earth and because pump filters can cause high earth leakage
current. Refer to
Note:
With respect to European Electromagnetic Compatibility (EMC) requirements for
harmonics and flicker, the iXM must be treated as Class A (industrial) as defined by
EN61326. The iXM is not intended for use in domestic buildings, or in properties directly
connected to an electrical supply network which also supplies domestic buildings.
All systems are supplied already configured for the electrical supply. Low voltage
(200 V a.c. to 230 V a.c.) and high voltage (380 V a.c. to 460 V a.c.) ranges cannot be
reconfigured.
Mains supply cable connection
WARNING: HAZARDOUS VOLTAGES
Risk of electric shock. The mains connector is not approved for connection and
disconnection under load.
Refer to
supply.
Two types of electrical supply connector are used on the system. Refer to
on page 18 for details of the connector types the systems to which they are connected.
M56635880_D
M56635880_D - Installation
Table: General electrical
Electrical supply safety
data.
on page 37 before you connect the system to the mains
Page 38
Electrical data
10/2020 - ©Edwards Limited