Fault
Warning 1.01 - Power interrupt
Cause
There has been a brownout of the electrical supply to the pump lasting more
than 1 second.
Remedy
Check the electrical supply.
Fault
Warning 31.13 - Gas missing
Cause
The gas module is not communicating with the pump controller.
Remedy
Contact us.
Fault
Warning 31.01 - SYS CONFIG FAULT
Cause
The wrong type of gas module has been installed.
Remedy
Contact us.
Fault
Warning 35.09 - N2Purge Low
Cause
The nitrogen supply flow rate is too low.
▪ The supply is possibly not connected or set to on.
▪ The supply line or filter is possibly blocked.
Remedy
Check the nitrogen supply is connected and set to on and that pipelines and filters
are not blocked.
Fault
Warning 35.11 - N2 Purge High
Cause
The Nitrogen supply flow rate is too high.
▪ The gas module and PDT settings do not match.
Remedy
Check the gas module and PDT settings, refer to
64.
Fault
Warning 35.13 - Gas Sensor Fault
Cause
The Nitrogen supply flow rate is too high.
▪ The gas module is possibly faulty or disconnected
Remedy
Contact us.
Fault
Warning 39.11 - Exh Press High
Cause
The pressure in the exhaust pipeline is too high.
▪ A valve in the pipeline is possibly shut.
▪ There is possibly process debris or condensation in the exhaust pipeline.
▪ You can have too many pumping systems connected to the exhaust pipeline.
M56635880_D