Figure 16 Protective Earth (PE) connection

Electrical connector locking mechanism
This product is approved permanently connected equipment and must be used as
described below:
The pumping system has an electrical connector locking mechanism
connectors on the rear of the
The pumping system is supplied with a protective cover
locking
following instructions to install the electrical supply cable.
Turn the locking screws to loosen the locking mechanism, if necessary.
Move the locking mechanism as far as possible to the left.
Push up the lever to release the protective cover.
Remove the protective cover and attach the electrical supply cable.
Pull the lever downwards to lock the electrical supply cable connector in place.
Push the locking mechanism as far as possible to the right so that it prevents the
lever from functioning.
Tighten the locking screws to firmly hold the locking mechanism in place.
Connect the other end of the electrical supply cable to the electrical supply
through an isolator.
M56635880_D
M56635880_D - Installation
mechanism) installed and the locking mechanism can be applied. Use the
system). Use an applicable screwdriver to release.