2 Installation and Wiring
Relationship between the Measurement Surface and Measured Dis-
tance
In order to measure the temperature more correctly, as shown in the figure below, install the infrared
thermal sensor at a distance where the measurement object can be captured as a large image, as far
as possible at the center of the field of view.
The relationship between the sensor visual field width and the distance from the infrared thermal sensor
to the measurement object is as shown below.
Sensor visual field width = 2 × (Distance from infrared thermal sensor up to the measurement object)
Example: When the distance from the infrared thermal sensor up to the measurement object is 100 mm
Sensor visual field width: 200 mm
Sensor visual field width: 200 mm
Viewing angle
Precautions for Correct Use
• Measurement objects:
Metals and transparent resin materials cannot be properly measured.
• Set the distance up to the measurement object in view of the occurrence voltage of the mea-
surement object and the safety certifications, etc.
2 - 6
Front view
Measurement
object
Field of view
per pixel: 6.25
mm angle
Top view
Measurement object
100 mm
90°
K6PM-TH Thermal Condition Monitoring Device User's Manual (H231)
Side view
Sensor visual field width: 200 mm
Viewing angle
90°
100 mm
Distance from
sensor up to the
measurement object
Distance from
sensor up to the
measurement object