4.1

External connections

Figure 4–1:
PDA Ethernet
connection
CM-A power
connection
ccManager
power
connection
SM-FL
power cord
4.2

Connecting Ethernet cables

The SM-FL includes an internal 10/100/1000 megabit Ethernet switch that accommodates the PC
(workstation) and as many as six ACQUITY UPC
cables from each module to the electronic connections on the rear panel of the SM-FL. The SM-FL
is connected internally to the Ethernet switch.
Requirement:
If you are running more than six modules in addition to the SM-FL, use a Waters
Ethernet switch box.
Core 1 configuration showing external cable connections
November 16, 2016, 715004521 Rev. C
Page 40
PDA power
connection
CO
inlet tube
2
connection
Inlet tube
bracket
Ethernet
switch
ccBSM power
connection
TP03459
2
system modules. Connect the shielded Ethernet