B.8

Connecting the workstation

Warning:
To avoid injury from electric shock or fire, and to prevent damage to the workstation and
ancillary equipment, do not place objects filled with liquid—such as solvent bottles—on these
items, or expose them to dripping or splashing liquids.
Before connecting the workstation to the instrument, set up the workstation according to its
accompanying instructions. Locate the workstation within 5 meters (16 feet) of the instrument.
Requirement:
Use shielded network cables with the instrument to reduce susceptibility to radio
and electrical frequency interference.
To connect the workstation:
1.
Connect the monitor to the PC.
2.
Connect one end of the shielded network cable to the appropriate port on the rear panel of
the instrument.
3.
Connect the other end of the shielded network cable to the port labeled Instrument LAN on
the workstation rear panel.
To connect the instrument to the power source:
Requirement:
installation procedures in the previous sections.
1.
Select the correct power cord for your location.
2.
Connect the female end of the power cord to the power port on the rear panel of the
instrument.
Do not connect the instrument's power supply cord until you complete the
September 7, 2016, 715003560IVD Rev. C
Page 137