Fitting the sampling cone assembly to the source
Required materials
Chemical-resistant, powder-free gloves
To fit the sampling cone assembly to the source:
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
the source enclosure open if one or both of these conditions apply:
• An ESI probe is fitted (the probe tip is sharp).
• A corona pin is fitted (the pin tip is sharp).
Caution:
fitting the sampling cone assembly to the ion block assembly.
The source components can be contaminated with
To avoid puncture wounds, take great care while working with
To avoid damage, do not open the source isolation valve before
Cleaning the sampling cone assembly
4-45