Fitting the sampling cone assembly to the source

Required materials
Chemical-resistant, powder-free gloves
To fit the sampling cone assembly to the source:
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
the source enclosure open if an ESI probe is fitted.
Caution:
fitting the sampling cone assembly to the ion block assembly.
1.
Ensure that the source isolation valve is in the closed position (see
page
4-10).
2.
Hold the sampling cone assembly so that the cone gas nozzle handle is
oriented horizontally and at the top, and then slide the sampling cone
assembly into the ion block assembly.
The source components can be contaminated with
To avoid puncture wounds, take great care while working with
To avoid damage, do not open the source isolation valve before
Cleaning the sampling cone assembly