To light the FID flame

To extinguish the FID flame

FID automatic reignition (Lit offset)

Advanced Operation Manual
LSV (height units)
-4
1.3 x10
-4
-4
pA
1.3 x 10
-3
-3
pA
4.2 x 10
-4
pA
device dependent
Press [Front Det] or [Back Det], scroll to Flame, then press
[On/Yes].
Press [Front Det] or [Back Det], scroll to Flame, then press
[Off/No].
Lit offset is the expected minimum difference between the FID
output with the flame lit and the output with the flame off.
The GC checks this value during runs and when loading a
method.
During a run, if the output falls below the Lit offset value, the
FID will attempt to reignite three times. If after the third
attempt the output does not increase by at least this value,
the detector shuts down all functions except temperature
and makeup gas flow.
When loading a method that includes a Flame On setting, the
GC performs a similar check. If the detector output is less
than the Lit offset, it will attempt reignition after reaching
method setpoints.
The default setting for Lit offset is 2.0 picoamps. This is a
good working value for all but very clean gases and systems.
You may want to lower this setpoint if the detector attempts
to reignite when the flame is still on, thus producing a
shutdown.
Area units
pA
1 pA-sec
-4
pA
1 x 10
pA
-3
pA
3.2 x 10
pA
-4
1.25 x 10
pA
5
Detectors
*
Noise (ASTM)
0.038 pA
0.038 pA
0.038 pA
0.038 pA
171