Setting Up Comparator Function; Turning On/Off Comparator Function; Setting Up Limit Range - Keysight Technologies E4981A Programming Manual

120 hz/1 khz/1 mhz capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Sorting Based on Measured Result (Comparator Function)

Setting Up Comparator Function

Setting Up Comparator Function

Turning ON/OFF comparator function

To enable/disable the comparator function, use the following command:

Setting up limit range

Clearing (resetting) limit range
You can clear the ON/OFF state and the lower and upper limit values of all limit ranges
(BIN1 to BIN9 and the secondary parameter limit range) and the ON/OFF state of AUX
BIN to recover the factory-shipped default values (refer to Operation Manual, Appendix C
" Initial Setting").
To clear the limit ranges, use the following command.
Selecting limit range designation method
You can select the designation method of the primary parameter limit ranges (BIN1 to
BIN9) from the following three modes:
NOTE
Only the absolute mode can be used as the designation method of the secondary parameter
limit range.
To select the designation method, use the following command.
To set up the reference value for the absolute tolerance mode or percent tolerance mode,
use the following command.
78
:CALCulate1:COMParator[:STATe] on page 187
:CALCulate1:COMParator:CLEar on page 176
Mode
Absolute mode
Absolute tolerance mode
Percent tolerance mode
*1.Boundary value - reference value
*2.((Boundary value - reference value)/reference value)∞100
:CALCulate1:COMParator:MODE on page 181
:CALCulate1:COMParator:PRIMary:NOMinal on page 184
Description
Designation using the absolute value
Designation using the
Designation using the absolute
relative value (deviation
value
from the reference value)
Designation using the percentage of
the reference value
*1
*2
Chapter 7

Advertisement

Table of Contents
loading

Table of Contents