[:Sense]:Correction:ckit:stan3[:Data] - Keysight Technologies E4981A Programming Manual

120 hz/1 khz/1 mhz capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

[:SENSe]:CORRection:CKIT:STAN3[:DATA]

[:SENSe]:CORRection:CKIT:STAN3[:DATA] <numeric 1>,<numeric 2>
Syntax
[:SENSe]:CORRection:CKIT:STAN3[:DATA]?
Description
Defines the values of the LOAD correction standard for the parameters you specify by
using the
These are set as the standard values for the measurement frequency when executing the
command (set with the
When using the multi-correction function (set to ON with the
[:SENSe]:CORRection:MULTiple[:STATe]
definition of the standard values enabled (set to ON with the
[:SENSe]:CORRection:MULTiple:CKIT:STAN3[:STATe]
for the selected channel (selected with the
command) are set at execution of the command.
Parameters
Description
Range
Initial value
Unit
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric 1},{numeric 2}<newline><^END>
Related commands
[:SENSe]:CORRection:LOAD[:STATe] on page 238
[:SENSe]:CORRection:CKIT:STAN3:FORMat on page 232
[:SENSe]:CORRection:MULTiple[:STATe] on page 241
[:SENSe]:CORRection:MULTiple:CHANnel on page 239
[:SENSe]:CORRection:MULTiple:CKIT:STAN3[:STATe] on page 240
Equivalent key
[
Meas Setup
sequence
Chapter 10
[:SENSe]:CORRection:CKIT:STAN3:FORMat
:SOURce:FREQuency[:CW]
<numeric 1>
Value of the primary
parameter.
-999.999 to 999.999
100E-9
F (farad)
] - CORRECTION - REF- A|B
Command Reference
E4981A SCPI Commands
command.
command).
command) with the channel-by-channel
command), the standard values
[:SENSe]:CORRection:MULTiple:CHANnel
<numeric 2>
Value of the secondary
parameter.
-99.9999E9 to 99.9999E9
0
Depends on the type of
parameter.
231

Advertisement

Table of Contents
loading