[:Sense][:Fimpedance]:Contact1:Verify:threshold1 - Keysight Technologies E4981A Programming Manual

120 hz/1 khz/1 mhz capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1

[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1 <numeric>
Syntax
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1?
Description
Sets/Gets the Threshold1 value for Contact check. Threshold1 is a parameter which is
proportional contact resistance of Hp or Hc. It is recommended to use the default (initial)
value.
NOTE
The available range for Hp, Hc and Lc is 220 μF to 1 mF at 120 Hz and 22 μF to 100 μF at
1 kHz. The contact resistance for Lp cannot be detected by contact check feature for this
range.
Parameters
Description
Range
Initial value
Resolution
Query response
<numeric><newline><^END>
Related commands
[:SENSe][:FIMPedance]:CONTact1:VERify[:STATe] on page 256
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2 on page 258
Equivalent key
[
Meas Setup
sequence
Chapter 10
Threshold1 value for Contact1.
0 to 1
0.1
0.01
] - CONT CHK- CC1 TH1
Command Reference
E4981A SCPI Commands
<numeric>
257

Advertisement

Table of Contents
loading

Table of Contents