Command Reference
E4981A SCPI Commands
[:SENSe]:DETector:DELay1
Syntax
[:SENSe]:DETector:DELay1 <numeric>
[:SENSe]:DETector:DELay1?
Description
Sets/Gets the waiting time for analog measurement for 120 Hz Frequency measurement.
Parameters
Description
Range
Initial value
Resolution
Query response
{numeric}<newline><^END>
Related commands
[:SENSe]:DETector:DELay2 on page 246
[:SENSe]:DETector:DELay3 on page 247
Equivalent key
No equivalent key can be used on front panel.
sequence
[:SENSe]:DETector:DELay2
Syntax
[:SENSe]:DETector:DELay2 <numeric>
[:SENSe]:DETector:DELay2?
Description
Sets/Gets the waiting time for analog measurement for 1 kHz Frequency measurement.
Parameters
Description
Range
Initial value
Resolution
Query response
{numeric}<newline><^END>
Related commands
[:SENSe]:DETector:DELay1 on page 246
[:SENSe]:DETector:DELay3 on page 247
Equivalent key
No equivalent key can be used on front panel.
sequence
246
<numeric 1>
Waiting time for 120 Hz measurement frequency
0 to 100m
1.67m
1u
<numeric 1>
Waiting time for 1 kHz measurement frequency
0 to 100m
1m
1u
Chapter 10