[:SENSe]:CORRection:COLLect[:ACQuire]

[:SENSe]:CORRection:COLLect[:ACQuire] {STAN1|STAN2|STAN3}
Syntax
Description
Measures the correction data for OPEN/SHORT/LOAD correction and turns on the
correction function.
For Open and Short correction, the correction data is measured for all measurement
frequencies. For load correction, the correction data is measured for the measurement
frequency used in executing the command (set with the
command).
When using the multi-correction function (set to ON with the
[:SENSe]:CORRection:MULTiple[:STATe]
data for the selected channel when executing the command (selected with the
[:SENSe]:CORRection:MULTiple:CHANnel
Parameters
STAN1
STAN2
STAN3
Related commands
[:SENSe]:CORRection:OPEN[:STATe] on page 244
[:SENSe]:CORRection:SHORt[:STATe] on page 245
[:SENSe]:CORRection:LOAD[:STATe] on page 238
[:SENSe]:CORRection:MULTiple[:STATe] on page 241
[:SENSe]:CORRection:MULTiple:CHANnel on page 239
Equivalent key
[
Meas Setup
sequence
[
Meas Setup
[
Meas Setup
Chapter 10
Specifies the OPEN correction.
Specifies the SHORT correction.
Specifies the LOAD correction.
] - CORRECTION - OPEN - MEAS OPEN
] - CORRECTION - SHORT - MEAS SHORT
] - CORRECTION - LOAD - MEAS LOAD
Command Reference
E4981A SCPI Commands
:SOURce:FREQuency[:CW]
command), this is measured as the correction
command). (No query)