Keysight Technologies E4990A Installation Manual

Keysight Technologies E4990A Installation Manual

Impedance analyzer
Hide thumbs Also See for E4990A:

Advertisement

Quick Links

Keysight E4990A
Impedance Analyzer
Installation Guide

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the E4990A and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Keysight Technologies E4990A

  • Page 1 Keysight E4990A Impedance Analyzer Installation Guide...
  • Page 3 Impedance Analyzer...
  • Page 4 Use, that is essential for the user to duplication or disclosure of Software is understand. subject to Keysight Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Gov- ernment will receive no greater than...
  • Page 5 When you notice any of the unusual conditions listed below, immediately terminate operation and disconnect the power cable. Contact your local Keysight Technologies sales representative or authorized service company for repair of the instrument. If you continue to operate without repairing the instrument, there is a potential fire or shock hazard to the operator.
  • Page 6 • The Low Voltage Directive 2006/95/EC • The EMC Directive 2004/108/EEC To obtain Declaration of Conformity, please contact your local Keysight Technologies sales office, agent or distributor. Safety Notice Supplement • This equipment complies with EN/IEC61010- 1:2001. • This equipment is of MEASUREMENT CATEGORY I (CAT I).
  • Page 7 N O T E POLLUTION DEGREE 2 in IEC61010-1. The E4990A is an INDOOR USE product. The LEDs in the E4990A are Class 1 in accordance with IEC60825-1, N O T E CLASS 1 LED PRODUCT. • Ground the Instrument To avoid electric shock, the instrument chassis and cabinet must be grounded with the supplied power cable’s...
  • Page 8 To avoid the danger of introducing additional hazards, do not install substitute parts or perform unauthorized modifications to the instrument. Return the instrument to a Keysight Technologies Sales and Service Office for service and repair to ensure that safety features are maintained in operational condition.
  • Page 9 Exclusive Remedies The remedies provided herein are Buyer’s sole and exclusive remedies. Keysight Technologies shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any other legal theory. Assistance...
  • Page 10 The installation guide (this manual) provides start up setup information when you use the E4990A for the first time and troubleshooting information when the Windows cannot be boot up. See this manual first when you use the E4990A for the first time. Online Help...
  • Page 11 The following shows the contents of this manual. Chapter 1, "Installation" This chapter provides information about how to set up the Keysight E4990A. Chapter 2, "Troubleshooting" This chapter describes the troubleshooting during start up and the procedure of the operating system (OS) recovery when the Windows OS has been damaged.
  • Page 12 Impedance Analyzer...
  • Page 13: Table Of Contents

    Verification and Connection of Power Cable Blown Fuses Starting the E4990A Turning the Power ON and OFF Disconnection from Supply Source Initial Registration of E4990A Troubleshooting Troubleshooting during Startup System Recovery Notes on executing the factory recovery function Procedure to execute the factory recovery...
  • Page 14 Impedance Analyzer...
  • Page 15: Installation

    Checking the Shipment 14 Environmental Requirements 15 Installing Front Handles/Rack Mounting Flanges 17 Power Supply 20 Blown Fuses 21 Starting the E4990A 22 Initial Registration of E4990A 25 This chapter provides information about how to set up the Keysight E4990A.
  • Page 16: Checking The Shipment

    Step 4. After checking, if one of the following applies, contact your nearest Keysight Technologies sales and service office. The packing box or shock- absorbing material used to package the analyzer has been damaged or the shock- absorbing material has traces where extreme force has been applied.
  • Page 17: Environmental Requirements

    For more information on environmental temperature to satisfy the specifications and measurement accuracy of the analyzer, see the specification in the E4990A Datasheet. When the environmental temperature around the analyzer is...
  • Page 18: Protection Against Electrostatic Discharge (Esd)

    AC outlet or the E4990A unit. When installing the E4990A, ensure that there is sufficient free space around the unit to permit quick disconnection of the plug (from AC outlet or E4990A unit) in case of emergency. Impedance Analyzer...
  • Page 19: Installing Front Handles/Rack Mounting Flanges

    Installation Installing Front Handles/Rack Mounting Flanges The E4990A can be installed on a workbench or on a rack. This section describes how to install the front handles (Option 1CN) used for moving or transporting the instrument, and how to install the analyzer in an equipment rack as part of a measurement system (Option 1CM: without the handles, Option 1CP: with the handles).
  • Page 20 Step 1. Remove the adhesive- backed trim strip (1) from each side of the outer frame of the E4990A front panel. Step 2. Use the provided screws to mount the •...
  • Page 21 Refer to Figure Step 3 (Option 1CM/Option 1CP). Remove the four bottom feet of the E4990A (lift the bar marked TAB on the inner side of the foot and slide the foot towards the bar). Step 4 (Option 1CM/Option 1CP).
  • Page 22: Power Supply

    Maximum power 300 VA consumption * 1. Switched automatically by the E4990A in conformity to the voltage used. Verification and Connection of Power Cable The three- wire power cable attached to the E4990A has one wire serving as the ground. Using this power cable allows the E4990A to be grounded, thereby protecting you against any potential electrical shock from the power outlet.
  • Page 23: Blown Fuses

    For any assistance, contact Keysight Technologies Customer contact center listed at the end of this guide. The E4990A uses the following fuse types: UL/CSA Type, Slow- Blo, 10 A- 250 Vac. DO NOT replace the fuse yourself as doing this may expose WA R N I N G yourself to electrical shock.
  • Page 24: Starting The E4990A

    Turning the Power ON and OFF The standby switch can turn ON/OFF the E4990A. The color on the button shows the status, as indicated below: Indicator Color...
  • Page 25: Disconnection From Supply Source

    Line switch (Always ON) and power cable receptacle Disconnection from Supply Source The power supply of the E4990A is cut off by disconnecting the plug of the power cable (on either AC outlet side or E4990A side). When it is necessary to disconnect the power supply in order to avoid shock hazards, etc., pull out the...
  • Page 26 Installation To allow this operation to be performed smoothly, be sure to follow the N O T E guidelines in Ensuring Adequate Free Space around Analyzer for Immediate Disconnection of Power Cable in Case of Emergency" on page When turning the power OFF under normal circumstances, always follow the methods described in Turning the Power OFF"...
  • Page 27: Initial Registration Of E4990A

    Installation Initial Registration of E4990A When you start up the E4990A for the first time, you need to perform the initial registration of the Windows operating system of the E4990A. When the Windows agreement message box appears, click Agree. The rest of the procedure runs automatically.
  • Page 28 Installation Impedance Analyzer...
  • Page 29: Troubleshooting

    Keysight E4990A Impedance Analyzer Installation Guide Troubleshooting Troubleshooting during Startup 28 System Recovery 29 This chapter describes the troubleshooting process during start up and the procedure of the operating system (OS) recovery when the Windows OS has been damaged.
  • Page 30: Troubleshooting During Startup

    Troubleshooting Troubleshooting during Startup When you encounter problems during start up, see Table System recovery saves most of problems. Table 1 Troubleshooting during startup Symptom Solution Turning ON (|) the standby switch does Confirm that the power cable is not start up the system. properly plugged in.
  • Page 31: System Recovery

    D drive) are not affected. However, Keysight recommends backing them up before executing system recovery for precautionary purposes. For more information on backup, refer to “Backing Up the Data” as described in E4990A Online Help. Procedure to execute the factory recovery This section describes how to return the contents of the C drive to the factory state.
  • Page 32 Step 3. Connect the keyboard to the E4990A. Step 4. Press the standby switch of the E4990A to turn it Step 5. When the screen as shown in the figure below appears, select Keysight Recovery System and press [Enter].
  • Page 33 Troubleshooting Step 6. Type 2 to select Recover Factory Backup Image option and press [Enter]. Figure 6 System recovery selection screen Step 7. A confirmation message appears. Click OK to continue the process. Figure 7 System recovery confirmation screen Impedance Analyzer...
  • Page 34 Initial Registration of E4990A" on page Never turn OFF the power during the system recovery because doing C A U T I O N so may cause serious damage to the E4990A. Impedance Analyzer...
  • Page 35: Index

    29 handle kit how to install the handle kit, 17 inspection checking the shipment, 14 Keysight Technologies sales and service office regional sales and support offices, 35 shipping consideration due to breakdown, etc., 6 manual Impedance Analyzer...
  • Page 36 Index Impedance Analyzer...
  • Page 37 Regional Sales and Support Offices For more information about Keysight Technologies test and measurement products, applications, services, and for the current sales office listing, visit our web site: http://www.keysight.com/find/tmdir. You can also contact one of the following centers and ask for a...
  • Page 38 Impedance Analyzer...
  • Page 39 This information is subject to change without notice. © Keysight Technologies 2014 Edition 2, March 2014 *E4990-90000* E4990-90000 www.keysight.com...