ASCII transfer format

When data is transferred in the ASCII transfer format, values are transferred as ASCII
bytes in one of the following formats. According to the specification of IEEE488.2, values
are separated by a comma (,).
NOTE
The string length of numerical values is not constant. Therefore, the comma does not occur
at constant intervals when extracting numerical values from the read out numerical data
string.
Integer format
The format shown in
example, the value, 11, is expressed as "+11."
Figure 6-4
Integer format
Floating point format
The format shown in
For example, the value, 1000, is expressed as "+1.00000E+03."
Figure 6-5
Floating point format
NOTE
The long format is available in floating point format. To set the long floating point format,
use the following command:
:FORMat:ASCii:LONG on page 219
Chapter 6
Figure
6-4. Numerical values are expressed as integers. For
Figure
6-5. Numerical values are expressed using a floating point.
Reading Out Measured Result
Data Transfer Format
67