[:SENSe]:CORRection:SHORt[:STATe]

[:SENSe]:CORRection:SHORt[:STATe] {ON|OFF|1|0}
Syntax
[:SENSe]:CORRection:SHORt[:STATe]?
Description
Turns ON/OFF the SHORT correction.
With the SHORT correction set to ON, if you change the cable length (set with the
:CALibration:CABLe[:LENGth]
:SYSTem:FSHift
Parameters
ON or 1
OFF or 0 (initial value)
Query response
{1|0}<newline><^END>
Related commands
:CALibration:CABLe[:LENGth] on page 174
:SYSTem:FSHift on page 276
[:SENSe]:CORRection:COLLect[:ACQuire] on page 233
Equivalent key
[
Meas Setup
sequence
Chapter 10
command) or frequency shift (set with the
command), the SHORT correction is automatically changed to OFF.
Turns ON the SHORT correction.
Turns OFF the SHORT correction.
] - CORRECTION - SHORT - ON|OFF