:CALCulate1:COMParator:SECondary:STATe

Syntax
:CALCulate1:COMParator:SECondary:STATe {ON|OFF|1|0}
:CALCulate1:COMParator:SECondary:STATe?
Description
Sets whether to enable sorting judgment for the measurement result of the secondary
parameter when using the comparator function.
Parameters
ON or 1
(initial value)
OFF or 0
Query response
{1|0}<newline><^END>
Related commands
:CALCulate1:COMParator:SECondary:LIMit on page 185
:CALCulate1:COMParator:PRIMary:BIN{1-9}:STATe on page 183
:CALCulate1:COMParator[:STATe] on page 187
Equivalent key
[
Meas Setup
sequence
186
Enables sorting judgment for the measurement result of the
secondary parameter.
Disables sorting judgment for the measurement result of the
secondary parameter.
] - LIMIT TABLE - BIN|2nd - ON|OFF