Command Reference
E4981A SCPI Commands
[:SENSe][:FIMPedance]:CONTact1:VERify:BUF2:FEED:C
ONTrol[:STATe]
Syntax

[:SENSe][:FIMPedance]:CONTact1:VERify:BUF2:FEED:CONTrol[:STATe]

{NEVer|ALWays}
[:SENSe][:FIMPedance]:CONTact:VERify:BUF2:FEED:CONTrol[:STATe]?
Description
Determines whether to get data buffer2 for Contact 1.
Parameters
ALWays
NEVer (Initial
value)
Query response
{NEV|ALW}<newline><^END>
Related commands
[:SENSe][:FIMPedance]:CONTact1:VERify:BUF2:DATA? on page 252
[:SENSe][:FIMPedance]:CONTact1:VERify:BUF2:FEED:CONTrol:INTerval on
page 253
on page 255
Equivalent key
No equivalent key is available on the front panel.
sequence
254
Feeds the contact check measurement data into the data buffer2
Does not feeds the contact check measurement data into the data
buffer2