[:Sense]:Correction:offset:data - Keysight Technologies E4981A Programming Manual

120 hz/1 khz/1 mhz capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

[:SENSe]:CORRection:OFFSet:DATA

Syntax
[:SENSe]:CORRection:OFFSet:DATA <numeric 1>,<numeric 2>
[:SENSe]:CORRection:OFFSet:DATA?
Description
Sets the correction values for the primary parameter and secondary parameter used in the
offset correction function.
These correction data are measured for the measurement frequency used in executing the
command (set with the
Parameters
Description
Range
Initial value
Unit
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
NOTE
The unit of parameter changes depending on the limit range designation method.
Query response
{numeric 1},{numeric 2}<newline><^END>
Related commands
[:SENSe]:CORRection:OFFSet[:STATe] on page 243
Equivalent key
[
Meas Setup
sequence
242
:SOURce:FREQuency[:CW]
<numeric 1>
The offset correction value for the
primary parameter.
-999.999 to 999.999
0
F (farad)
] - CORRECTION - OFFSET - A|B
command).
<numeric 2>
The offset correction value for the
secondary parameter.
-99.9999E9 to 99.9999E9
0
Depends on the setup of the
secondary parameter.