Command Reference
E4981A SCPI Commands

:TRIGger[:SEQ1][:IMMediate]

Syntax
:TRIGger[:SEQ1][:IMMediate]
Description
Immediately generates a trigger and executes a measurement. This command is valid for
trigger source values of MANual or BUS trigger specified by
command. On setting the trigger source value to INTernal or EXTernal, this command
generates an error.
If the trigger system is not in the trigger wait state (trigger event detection state), executing
this command causes an error and the command is ignored.
For details on the trigger system, refer to
Equivalent key
[
Trigger
sequence

:TRIGger[:SEQ1]:SLOPe

Syntax
:TRIGger[:SEQ1]:SLOPe {POSitive|NEGative}
:TRIGger[:SEQ1]:SLOPe?
Description
Determines the trigger polarity for BNC external trigger on the rear panel. This does not
affects the trigger signal on handler and scanner interface.
Related commands
*TRG on page 169
Parameters
POSitive (Initial
[Factory set] value)
NEGative
NOTE
The value of this command is not changed with
changed with Factory Default reset (available only through Front Panel).
Query response
{POS|NEG}<newline><^END>
Equivalent key
[
System
sequence
298
]
Detects positive edge trigger signal.
Detects negative edge trigger signal.
] - EXT TRIG POL - POS|NEG
:TRIGger[:SEQ1]:SOURce
"Trigger system" on page
Description
*RST
and
:SYSTem:PRESet
60. (No query)