:DISPlay[:WINDow]:TEXT1[:DATA]:FMSD:DATA

:DISPlay[:WINDow]:TEXT1[:DATA]:FMSD:DATA <numeric>
Syntax
:DISPlay[:WINDow]:TEXT1[:DATA]:FMSD:DATA?
Description
Sets the value of the highest digit of the primary parameters in the fixed point display. The
parameter is always fixed to the same value, regardless of this setting, if the following
conditions are met:
Deviation is displayed as a percentage in the deviation measurement mode.
The resulting D, Q and % of the secondary parameter is displayed.
Parameters
Description
Range
Initial value
If one of the settable values is not specified for the parameter, the minimum possible value,
which is larger than the specified parameter value, is set. If the specified parameter exceeds
the maximum value, the maximum value is set.
MAX or MIN can be used to specify the parameters.
Query response
{numeric}<newline><^END>
Related commands
:DISPlay[:WINDow]:TEXT1[:DATA]:FMSD[:STATe] on page 214
:CALCulate1:MATH:EXPRession:NAME on page 189
:CALCulate1:MATH:STATe on page 190
:CALCulate2:MATH:EXPRession:NAME on page 192
:CALCulate2:MATH:STATe on page 193
Equivalent key
[
Display Format
sequence
POS DECL-
Chapter 10
The value of the highest digit of the primary parameter.
1a|10a|100a|1f|10f|100f|1p|10p|100p|1n|10n|100n|1u|10u|100u|1
m|10m|100m|1|10|100|1k|10k|100k|1M|10M|100M|1G|10G|100
G|1T|10T|100T|1P|10P|100P|1E
1n
] - <Primary Measurement Parameter i.e Cp/Cs> - D.P. POS INCR+|D.P.
Command Reference
E4981A SCPI Commands
<numeric>
213