Calculate2:Format; Calculate2:Math:expression:catalog - Keysight Technologies E4981A Programming Manual

120 hz/1 khz/1 mhz capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

:CALCulate2:FORMat

:CALCulate2:FORMat {D|Q|G|RP|RS}
Syntax
:CALCulate2:FORMat?
Description
Specifies the secondary parameter to be measured.
If the primary parameter has been set to CP and the secondary parameter is set to RS, the
primary parameter is automatically changed to CS. Also, if the primary parameter has been
set to CS and the secondary parameter is set to G or RP, the primary parameter is
automatically set to CP.
Parameters
D (initial value)
Q
G
RP
RS
Query response
{D|Q|G|RP|RS}<newline><^END>
Related commands
:CALCulate1:FORMat on page 188
Equivalent key
[
Meas Setup
sequence

:CALCulate2:MATH:EXPRession:CATalog?

Syntax
:CALCulate2:MATH:EXPRession:CATalog?
Description
Reads out available parameters used when executing the command to specify the
expression of the secondary parameter in the deviation measurement mode (the
:CALCulate1:MATH:EXPRession:NAME
DEV,PCNT. (Query only)
Query response
DEV,PCNT<newline><^END> (fixed)
Related commands
:CALCulate2:MATH:EXPRession:NAME on page 192
Equivalent key
No equivalent key is available on the front panel.
sequence
Chapter 10
Specifies the dissipation factor as the secondary parameter.
Specifies the quality factor (inverse value of D) as the secondary
parameter.
Specifies the equivalent parallel conductance measured using the
parallel equivalent circuit model as the secondary parameter.
Specifies the equivalent parallel resistance measured using the
parallel equivalent circuit model as the secondary parameter.
Specifies the equivalent series resistance measured using the series
equivalent circuit model as the secondary parameter.
] - FUNC - Cp{D|Q|G|Rp}|Cs{D|Q|Rs}
Command Reference
E4981A SCPI Commands
Description
command). The query response is always
191

Advertisement

Table of Contents
loading

Table of Contents