Trend Monitor - Omron Z500 Operation Manual

Profile measuring system
Hide thumbs Also See for Z500:
Table of Contents

Advertisement

SECTION 2
Basic Operations

2-3-4 Trend Monitor

Mostly used for setting and checking the Trigger function.
By using the Trend Monitor, the measurement results to be monitored can be checked in chronological
order for up to 30 seconds. Press the TRIG Key to switch from waveform display to waveform display
stop and vise versa. The trigger function enables waveform to be displayed at specified timing.
Scn 0
Thru
OUT 0
+0005.18800
S+
/
Output number
The current output number is displayed.
Screen Mode
Thru: The waveform obtained by the Sensor is displayed continuously.
Trig: A still image of the waveform based on the trigger settings is displayed.
Thin lines are used to represent the measurement start and measurement end timings.
CHECK
CHECK
Z500
36
Operation Manual
Conversion values
for X and Y axes
Run
9.96ms
1div x: 1.000 [s]
y:+10.0000[mm]
Zero's OFF
mm
PASS
:Display S+ENT:Adjust
Measurement
Judgment result
value
Display range and display time can be changed. (Only the display range can be changed
for each output number.)
REFERENCE
Refer to page 118.
Characteristic quantities (such as average value and peak value being held) at specified
timing can be output.
REFERENCE
Refer to page 84.
The units corresponding to 1 division (1 div) on the
waveform display are displayed. If the setting is
changed, the display also changes. Use these
values as reference when displaying the waveform.
y
1div x:
1div
x
Measurement period (Sampling time)
Waveform display status
During waveform display,
While the waveform is stopped,
Trigger waveform
The trigger status is displayed.
Force-zero
Zero's OFF: Force-zero not executed
Zero's ON: Force-zero executed
1.000
y:+10.000
(green) is displayed.
(red) is displayed.