Step 6 Output Settings - Omron Z500 Operation Manual

Profile measuring system
Hide thumbs Also See for Z500:
Table of Contents

Advertisement

SECTION 4
Menus for Expert Menu
6
Output Settings
Step
The Z500 can transmit data to an external device in three formats; analog, terminal, and RS-232C. In
the example provided below, the procedure for setting the output contents to the external device is
explained. The output data and their transmission methods are as follows:
Data
Measure-
The measurement results for each output number are output
ment value
in numerical values.
Judgment
The judgment results for each output number are output.
result
Data on height of 126 lines are output.
Profile
When setting to Analog, analog waveforms are output.
Only the Trigger function allows this type of output.
Trend
Data that have been held for each output number are output.
Analog
Outputting Measurement Values
Item
Selection/Setting
OUT0 to OUT7
Output
Profile
(OUT0*)
-9999.99999
4mA (-5V)
to 9999.99999mm
-9999.99999
20mA (+5V)
to 9999.99999mm
Z500
100
Operation Manual
Description
Measured values or profile data are converted to analog values and are output.
Setting items vary depending on the outputting data.
Use the procedure explained below to set the measurement values correspond-
ing to the upper limit (20 mA/5 V) and lower limit (4 mA/-5 V) for analog output.
This setting enables output to be scaled (e.g., 1 mm corresponds to 1 mA) as
required. Set both the analog output and measurement values so that the upper
limit is greater than the lower limit.
Analog
20mA
(+5V)
4mA
(-5V)
Measurement
Selects the output number to be set.
Sets the lower limit for the analog output measurement range.
The default setting varies depending on the Sensor model connected.
Sets the upper limit for the analog output measurement range.
The default setting varies depending on the Sensor model connected.
Analog
p.100
supported
Not
supported
supported
Not
supported
Measurement value
Measurement
lower limit
upper limit
Description
The asterisk (*) indicates the default setting.
Terminals
RS-232C
p.103
p.105
Not
supported
supported
supported
supported
Not
supported
supported
Not
supported
supported