Environment Settings - Omron Z500 Operation Manual

Profile measuring system
Hide thumbs Also See for Z500:
Table of Contents

Advertisement

SECTION 6
System Settings

6-3 Environment Settings

Environment settings for the Z500's system can be made using the items shown below.
Item
Settings
Sensor
Scene
Monitor
1st scene
Dec. pts
Download
Z500
146
Operation Manual
Selection
Conversatn*
Expert
Diffuse*
Mirror
Console*
Terminals
Image*
Digital
Trend
Profile
Scn 0* to Scn 15
1 to 5* digits
ON
OFF*
Description
Select the menu used to set the measurement
conditions.
Select the Sensor mounting method. Select the item that
corresponds to the mounting method used for the Sensor
connected. The default setting varies depending on the
Sensor model connected.
Select whether scene changing is performed on the
Console or using the terminal block.
Select the type of screen that is displayed when the
power supply is turned ON.
Select the scene number that is displayed when the
power supply is turned ON.
Select the lowest digit of measurement values displayed
on the monitor.
The Z500 acquires information from the sensor at star-
tup. The startup time can be shortened by downloading
this information into the controller and storing it.
•ON
The information stored inside the controller is acquired
at startup. The startup time will be shortened by compar-
ison to when the OFF function is used.
•OFF
Information is acquired via the sensor connection at
startup. The startup time will be longer than when the
ON function is used.
NOTICE
To replace the sensor for connection to the controller,
perform the following procedure:
1. Turn OFF the download function.
2. Replace the sensor.
3. Download the new sensor information.
4. Turn ON the download function.
If the sensor is replaced when the download function is
ON, measurements cannot be taken correctly with the
new sensor, because the old sensor information (such as
the measurement range) is still stored.
Also, perform the above procedure when a sensor of the
same model is used to replace the existing sensor.
The asterisk (*) indicates the default setting.

Advertisement

Table of Contents

Troubleshooting

loading

This manual is also suitable for:

Z500-mc15eZ500-mc10e