Compensate the Measured Level Difference and Width
To compensate for the error, perform the measurement once.
Applicable measurement items: Step: 2 pts, Step: 3 pts, Width
Example : When selecting Step: 2 pts
Level difference: 5 mm
Use the method shown below to calibrate workpieces so that a displacement value of 4.95 mm is
measured as 5 mm.
Point 2
To compensate for this error, perform the measurement once.
Measure with the
workpiece in position.
1.
Move the cursor to Step: 2 pts and
press the ENT Key.
The selections will be displayed.
2.
Select Calibrate.
5mm
4.95mm
Calibrate to 5 mm.
Point 1
Input the value
of point 1.
0mm
Measurement value
Compensate for this error.
Level difference of workpiece
5mm
Input the value
of point 2.
5mm
Scn 0
Set
OUT0:Height
OUT1:Step: 2 pts
OUT2:Edge position
OUT3:Define
OUT4:
OUT5:
OUT6:
OUT7:
Scn 0
Set
Change
Calibrate
Output
Delete
SECTION 4
Menus for Expert Menu
Z500
Operation Manual
81